7-58
TSL-710 Operation Manual
TM Syntax TM<value>
Description Setsthetimingofthetriggersignaloutput.Whenparameterisnot
specied,readsoutthecurrentlysetvalue.
Parameter 0:None
1:Stop
2:Start
3:Step
Response Currently set value
Example SetsthetimingtoStop.
TM1
TW Syntax TW<value>
Description Setstheintervalofthetriggersignaloutput.Whenparameterisnot
specied,readsoutthecurrentlysetvalue.
Parameter Range:0.0001~160(nm)
Step: 0.0001(nm)
Response Currentlysetvalue.Avalueisreturnedbothonsettingandonreading.
Example Setstheintervalofthetriggersignaloutputto1nm.
TW1
CO Syntax CO
Description TurnsOnthecoherencecontrol.
CoherencecontrolcannotbechangedtoON.whenmodulation
settingis“Ext.”.
Parameter None
Response None
CF Syntax CF
Description TurnsOFFthecoherencecontrol.
Parameter None
Response None
GA Syntax GA<value>
Description SetsGPIBaddress.Whenparameterisnotspecied,readsoutthe
currentlysetvalue.
Parameter Range: 1~30
Step: 1
Response Currently set value
Example SetsGPIBaddressto20.
GA20