Tunable Semiconductor Laser TSL-710
7-61
Detaildescriptionofthe4thdigit
TN Syntax TN
Description Readsthenumberofdatapointsrecordedbywavelengthlogging.
Parameter None
Response 00000~65535(Thenumericalvalueofvedigits)
Example TN
01000
Indicatesthatrecordeddatapointsare1000.
TD Syntax TD
Description Readsthedatarecordedbywavelengthloggingonedatarecordata
time.ThedataisreturnedinASCIIcharacters.
Thereadingpointmovestothenextdataafterreadingonedata.
Parameter None
Response Wavelengthdatainnm.
Example TD
1555.000(ASCIIcharacters)
Recordeddataatthispointis1555nm.
TA Syntax TA
Description Readsoutalldatarecordedbywavelengthloggingatatime.Adatais
in4-bytebinary,andtransmittedinBigendians.Dataformatis
integernumberin0.1pmunits.
Parameter None
Response Binarydataarray
Example TA
00E6686000E68F7000E6B680(inHexadecimalnumber)
Itincludesthree4-bytedataandtheyaretransformedto
151000001511000015120000indecimalnumbers.
Value
Power control Attenuator control
Power monitor
range control
0 Auto Hold Auto
1 Manual Hold (Manual) Auto
2 Auto Auto Auto
4 Auto Hold Hold
5 Manual Hold (Manual) Hold