7-60
TSL-710 Operation Manual
-□□□□□□
Code
6
th
digit
5
th
digit
4
th
digit
3
rd
digit
2
nd
digit
1
st
digit
SU Syntax SU
Description Readsoutthestatus.
Parameter None
Response Status.Statusisshownbytherstsignandthenumericalvalueofsix
digits.Thestatethateachdigitshowsisasfollows.
Code[-/none]indicatesLDstatus.0:OFF,1:ON
6thdigit[0/1]indicatescoherencecontrol.0:OFF,1:ON
5thdigit[0/1]indicatesFine-tuning.0:OFF,1:ON
4thdigit[0-5]indicatescontrolmodeofoutputpower,attenuator,and
powermonitorrange.Fordetailrefertothetablebelow.
3rddigit[0/1]indicatesLDtemperatureerror.
0:Noerror,1:Erroroccurred.
2nddigit[0/1]indicatesLDcurrentlimiterror.
0:Noerror,1:Erroroccurred.
1stdigit[0-7]indicatesoperationstatus.
0:Operationiscompleted.
1:Wavelengthistuning.
2:LDcurrentissetting(LDisonstateandpowercontrolisAuto).
3:WavelengthistuningandLDcurrentissetting.
4:Attenuatorissetting.
5:wavelengthissettingandattenuatorissetting.
6:LDcurrentissettingandattenuatorissetting.
7:Wavelengthistuning,LDcurrentissettingandattenuatorissetting.
Example Readsstatus.
SU
→-000005LDisON,wavelengthtuningisoperatingand
attenuatortuningisoperating.