Tunable Semiconductor Laser TSL-710
7-49
2. TSL-710 specic command detailed description
Specic commands in Command set 2 are described in detail as follows.
*IDN? Syntax *IDN?
Description A query to identify the device; places strings of device information
suchasthename,rmware,andversionintheoutputqueue.
Parameter None
Response SANTEC,TSL-710,########,****.****
#eld=serialnumberofthedevicein8digits.
*eld=rmwareversionas4digits+.(period)+4digits.
Example *IDN?
→ SANTEC,TSL-710,06020001,0001.0000
LO Syntax LO
Description SetsONtheLD.
Parameter None
Response None
LF Syntax LF
Description SetsOFFtheLD.
Parameter None
Response None
WA Syntax WA<value>
Description Setstheoutputwavelength.Whenparameterisnotspecied,reads
outthewavelengthcurrentlyset.
Parameter Range: MinimumsettingwavelengthtoMaximum(nm)
Step: 0.0001(nm)
Response Currentlysetvalue.Avalueisreturnedbothonsettingandonreading.
Example Setsoutputwavelengthto1560.123nm.
WA1560.123
Reads out the currently set value
WA
→ 1560.000
Indicatesthatthecurrentlysetwavelengthis1560nm.
FQ Syntax FQ<value>
Description Setstheoutputfrequency.AsetunitisTHz.
Whenparameterisnotspecied,readsoutthefrequencycurrentlyset.
Parameter Range: MinimumsettingfrequencytoMaximum(THz)
Step: 0.00001(THz)