7-50
TSL-710 Operation Manual
Response Currentlysetvalue.Avalueisreturnedbothonsettingandonreading.
Example Setsoutputfrequencyto192.0000THz.
FQ192.0000
Reads out the currently set value
FQ
192.0000
Indicatesthatthecurrentlysetfrequencyis192.0000THz.
FT Syntax FT<value>
Description SetstheFine-Tuningvalue.
Whenparameterisnotspecied,readsoutthevaluecurrentlyset.
Parameter Range: -100~+100
Step: 0.01
Response Currentlysetvalue.Avalueisreturnedbothonsettingandonreading.
Example SetstheFine-Tuningvalueto–50.
FT-50
Reads out the currently set value
FT
0100.00
Indicatesthatthecurrentlysetvalueis100.
FTF Syntax FTF
Description StopsFine-tuningmode,andstartsclosed-loopwavelengthcontrolling.
Paramete None
Response None
OP Syntax OP<value>
Description SetstheopticaloutputpowerindBm.
Whenparameterisnotspecied,readsouttheactualpowervalue.
Parameter Range: -20~+10dBm(typical)
Step: 0.01(dB)
Response Currentlysetvalue.Avalueisreturnedbothonsettingandonreading.
Example Setstheopticaloutputpowerto+3dBm.
OP3
Readsouttheactualpowervalue.
OP
002.995
Indicatesthattheactualpowervalueis2.995dBm.