7-20
TSL-710 Operation Manual
3.
IEEE-488.2 common commands - Detailed description
1) *IDN? Identication Query
Syntax *IDN?
Description A query to identify the device; places strings of device
information such as the name, rmware, and version in the
output queue.
Parameter None
Response SANTEC,TSL-710,########,****.****
# eld = serial number of the device in 8 digits.
* eld = rmware version as 4 digits + .(period) + 4 digits.
Example *IDN?
SANTEC,TSL-710,06020001,0001.0000
2) *RST Device Reset
Syntax *RST
Description Aborts standby operation.
Clears the following items.
Command input queue
Error queue
Parameter None
Response None
3) *TST? Self-test Query
Syntax *TST?
Description Initiates an instrument self-test and places the results in the
output queue. It performs internal calibration of laser
and/or diagnostic.
Initializes motors, checks output power, LD current,
and LD temperature.
It takes a couple of minutes.
Response This should return “0” for success, non-zero return values
for error conditions. Return value is a combination of the
following numbers.
0: No error
1: LD temperature is out of range (at LD OFF)
2: LD temperature is out of range (at LD ON)
4: Wavelength monitor temperature is out of range
8: LD injection current is overload
16: Power monitor is malfunction

Table of Contents