7-30
TSL-710 Operation Manual
:TRIGger:OUTPut
Syntax :TRIGger:OUTPut <wsp> <value>
Description Sets the timing of the trigger signal output.
Parameter  0None
1Stop
2Start
3Step
Response None
Example :TRIG: OUTP 1
Sets the timing of the trigger signal output to be at
sweep completion.
:TRIGger:OUTPut?
Syntax :TRIGger:OUTPut?
Description  Reads out the timing setting of the trigger signal output.
Parameter None
Response 0None
1Stop
2Start
3Step
Example :TRIG:OUTP?
2
Indicates that the timing of the trigger signal output
is at sweep start.
:TRIGger:OUTPut:STEP[:WIDTh]
Syntax :TRIGger:OUTPut:STEP[:WIDTh]<wsp><value>
Description Sets the interval of the trigger signal output.
Parameter  Range0.0001 160 ( nm     
Step0.0001nm
Response None
Example :TRIG:OUTP:STEP 1
Sets the interval of the trigger signal output.
:TRIGger:OUTPut:STEP[:WIDTh]?
Syntax :TRIGger: OUTPut: STEP [:WIDTh]?
Description  Reads out the interval of the trigger signal output.
Parameter None
Response Range0.0001 160 ( nm
Step0.0001nm
Example :TRIG:OUTP: STEP?
10.000
Indicates that the trigger signal output interval is 10 nm.