110 BioPAT
®
Trace | Multi Trace Operating Instructions
Operation
Status Action LED1
(green)
LED 2
color
LED 2
status
Calibrations
Calibration loaded on blue on
Calibration unloaded on blue flashes
Reference loaded on blue on
Reference unloaded on blue flashes
Malfunctions
No network
connection
on red on
Initialisation error on black off
8.5 Tips
8.5.1 Choosing Suitable Standard Solutions
Various combinations of calibration standard solutions are available to
optimally cover the expected concentration cycles. Sartorius recommends
connecting a higher concentrated standard than standard1.
Depending on the process conditions and the objective of the cultivation,
various concentration ranges have to be covered. The concentration range
should principally be between the concentrations of the standard solutions in
order to ensure optimal measuring accuracy. Furthermore, we recommend
always choosing a standard with very low concentration in order to properly
determine measured values close to zero.
For regulation, the concentrations of the calibration standards should not be
too far from the regulation (set)point. For regulation at 1g/L, for example, it is
more practical to use a combination of 0.5g/L and 4g/L than a combination
of 1g/L and 10g/L, for example.