BioPAT
®
 Trace | Multi Trace Operating Instructions   87
  Operation 
By clicking on the button “Continue measure-
ment”, the system switches directly to opera-
tional mode. No rinsing cycles or calibrations will 
be carried out. The former calibration values will 
be retained.
NOTICE
Do not activate “Continue measurement” unless 
you are certain that the same conditions prevail 
as before the pause (e.g. in the event of brief 
interruptions). Otherwise, this may result in 
erroneous measurements!
In the case of longer interruptions (e.g. over-
night), it is recommended to carry out a calibra-
tion before continuing with further 
measurements.
8.3.2  Menu Item “Display”
When the menu item “Show Connection” is 
activated, an asterisk indicates that the control 
PC is communicating properly with the BioPAT
®
 
Trace | Multi Trace. 
8.3.3  Print Diagram
By selecting the menu item “Print diagram”, the 
currently displayed diagram is printed out on an 
interfaced printer.