46 BioPAT
®
Trace | Multi Trace Operating Instructions
Initial Installation
6.1.3 Attaching the Tubing Set for the
“Dialysis” Operating Mode
Before attaching the filtration tubing set, the
valve tubes must first be inserted into the pinch
valves. To keep them from getting mixed up,
each line is color-coded.
The following rules apply:
— The pair of tubes are color-coded (red, yellow
and green) and are kept in place by two
guidance plates
— The top-to-bottom order follows the traffic
light principle: red at the top, yellow in the
middle, green at the bottom
— The color-coded line of the pair of tubes is
always inserted into the front valve slot
— When inserting the tubing, always guide the
labeled side with your right hand
After you have inserted the tubing set, check that
the result matches this picture.
As the first step, take both guidance plates in
both hands and open them slightly in order to
align the valve tubes in parallel.
The pair of tubes marked in red should be on top
and the label on the right should be visible.