122 BioPAT
®
Trace | Multi Trace Operating Instructions
Malfunctions
10.2 Error and Possible Causes
Error | Error message Possible cause
Calibration failed Calibration standard empty
Transport buffer empty
Air bubble inside the measuring chamber
Incorrect tubing set used
Conductivity cell not properly inserted
Wrong method selected
Valve tubes inserted incorrectly or not
inserted
Dialysis only: three-way stopcock incor-
rectly adjusted
Tubing set out of date
Leaks in the tubing set
Calibration line blocked or tubing set
clogged
Software versions of device | PC not
compatible
Valve (yellow) cannot be triggered
Measured value too high
or too low
Calibration erroneous
Parameter entry
“Standard” parameter entry and actual
standard do not match
Air bubble inside the measuring chamber
Transport buffer empty
Leaks in the tubing set
Dialysis only: Reference measurement not
performed
Dialysis only: Reference measurement not
up-to-date