BioPAT
®
 Trace | Multi Trace Operating Instructions   33
    Design and Function 
4.3  Conductivity Cell Function
The conductivity cell contains biosensors for 
glucose and lactate or methanol/ethanol. The 
conductivity cell is shown in the figure on the left. 
The fluid inlet flow runs perpendicular to the 
2-channel sensor (“wall jet” principle). It is 
possible to check that the flow through the cells 
is free from air bubbles by using the viewing 
window. Any air bubbles can be removed as 
necessary by rinsing the measuring chamber. 
Forthat purpose, the fluid is injected through the 
rinsing line equipped with a return valve.
Contact springs intercept the electric signal 
measurable at the electrodes and transmit it to 
the BioPAT
®
 Trace | Multi Trace. The conductivity 
cell is fastened to the device with a knurled nut.
 CAUTION
Inaccurate measurement result due to air bubbles!
It must be ensured that there are no air bubbles in the measuring chamber of 
the flow cell, as they can distort the results.
After connecting the sampling probe and before starting each measurement 
series, check the conductivity cell for air bubbles through the viewing window.
As a general rule, air bubbles in the tubing set are flushed out over time. 
However, this process takes a certain amount of time (e.g. during the priming 
process).