Trace | Multi Trace Operating Instructions
Malfunctions
10.1.1 Calibration Error
For each calibration, the resulting measurements must lie within given toler-
ances, otherwise this error message appears. Both tables (see “10.2 Error and
Possible Causes”, page 122 | “10.3 Error Message: “Calibration Failed””,
page 124 give an overview of possible causes and repair.
10.1.2 Calibration Implausible
This test is a plausibility check, in which, although the calibration is running
smoothly, false results occur (for instance if the standard concentrations are
typed in wrong). For each calibration, the resulting values (offset | slope) must
lie within given ranges, otherwise this error message appears. Both tables (see
“10.2 Error and Possible Causes”, page 122 | “10.3 Error Message: “Calibra-
tion Failed””, page 124 give an overview of possible causes and repair.