BioPAT
®
 Trace | Multi Trace Operating Instructions   83
  Operation 
The “Fast Reference” option allows a faster 
implementation of the reference measurement, 
in which the known concentration values of the 
medium can be entered before the actual 
reference measurement. After pressing the “Fast 
Reference” button, the input window appears.
Now enter the corresponding concentration 
values and confirm with “Apply”. The device will 
then begin with the reference measurement and 
when finished automatically returns to the state it 
was in when the “Fast Reference” button was 
triggered. “Reject” will cancel the operation and 
return to the starting position.
As long as the input window is open and no new 
reference values have been registered and 
confirmed, the BioPAT
®
 Trace | Multi Trace 
Control still works with the current reference 
factor.
8.3  Menu Bar
The measured data are recorded as text in a bpd 
file and saved in the default subdirectory “Data”. 
In addition to the date and time the respective 
data for temperature at the measurement cell, 
glucose reading and lactate reading will be 
logged. If the controller is active, the respective 
feed rate (1 = 100%) is recorded. If the device is in 
the Multi Trace mode, the port for which the 
values apply is logged as well.