 Malfunctions
10.3  Error Message: “Calibration Failed”
Possible cause What should you do?
1. The transport buffer is empty (Is there 
still enough transport buffer in the 
storage vessel? Is the suction tubing 
still in the storage vessel? Is buffer 
dripping into the drip counter?)
Connect new transport 
buffer and fill the tubing 
set.
2. A calibration solution is empty. Connect new calibration 
solution.
3. Silicone tubes are not properly inserted 
into the valves.
Check if the valve tubes 
are seated properly and 
correct as required.
4. There is an air bubble inside the 
measuring chamber.
Restart the device and fill 
again.
5. A tube is pinched (e.g. when screwing 
the covers into the standard bottles.)
Remove pinch.
6. There is clogging in the tubing set 
(When the buffer remains in the tubing 
set over an extended period of time, 
salts begin to crystallize out. 
Crystallization usually occurs on the 
silicone tubes in the valves. Clogging 
can also occur due to microbial 
growth.)
Remove clogging. If this is 
not possible, use a new 
tubing set.
7. Leaks (If the clog is not removed, tubes 
may burst.)
Correct leaks. If this is not 
possible, use a new tubing 
set.
8.  Sensor does not have proper contact 
(When the sensor is not inserted 
properly, the reverse side of the sensor 
does not have proper electrical contact 
with the sensor holder.)
Remove sensor and 
reinstall it properly.