BioPAT
®
Trace | Multi Trace Operating Instructions 123
Malfunctions
Error | Error message Possible cause
No or negative measured
value
Conductivity cell not properly inserted
Transport buffer empty
Tubing set clogged
Leaks in the tubing set
Tubing set out of date
No sampling probe or wrong sampling
probe connected
Measured value varies
greatly
Air bubble inside the measuring chamber
Conductivity cell not properly inserted
Tubing set out of date
No connection to
trace_mon
IP address set incorrectly
PC: IP address not automatically assigned
Incorrect or defective power cord
Device boots but program does not start
Pump is not turning Remove pump cassette and clean
motorshaft
Pull tight pump tube
No flow Transport buffer empty
Calibration standard empty
Tubing set clogged
Leaks in the tubing set