66 BioPAT
®
 Trace | Multi Trace Operating Instructions
  Initial Installation
6.4.1  Troubleshooting
If you have entered different or incorrect IP addresses on the device website 
and during the installation, it will not be possible to establish a connection. 
The following options are available:
t First, attempt to produce a connection to the internal device website via 
the standard IP address 192.168.120.231 using a direct PC connection and 
a cross-over Ethernet cable. 
t If this is not possible, a change to a known IP address must be forced using 
the boot file. First switch the device off. Then copy the boot file “ip_set” 
onto a USB stick and connect it to the rear of the device. Then switch the 
device on. Once the boot process is complete, the device will now have 
the new IP address 192.168.120.232. 
t Navigate to the new IP address 192.168.120.232 in your web browser.
y After the device has been connected in this manner, you can set the 
network address via your web browser as described above.
6.5  Installing the Software
t Run the file “setup_tracemon” to install the operating software on your 
computer. When doing so, you will be asked to enter the new IP address.
t  In order to do this, overwrite the suggested factory-set IP address. If you 
have not changed the IP address then use “Next” to confirm.
t During the installation you will be asked to specify the storage locations 
for the program and the subsequent measurement data. 
y Next, the software will start the device with the specified IP address and 
automatically connect to the BioPAT
®
 Trace | Multi Trace.