    Maintenance and Servicing 
9.2.2  Causes of Contamination
 — Connection lines: The connection lines are neither cleaned nor disinfected 
prior to use.
 — Transport buffer: The transport buffer is prepared in a contaminated 
container or filled into an already used standard container.
 — Tubing set: The tubing set is used again without having been thoroughly 
cleaned or disinfected beforehand.
 — Disposal line | disposal tank: The disposal line is filled with liquid. 
Nodisinfectant was added to the disposal tank. The disposal tank has not 
been closed.
 — Probe sterilization: The probe has not been properly sterilized or 
contaminated with a rinse solution.
9.2.3  Decontaminating Connection Lines
t Rinse the connection lines with a cleaning solution.
t Rinse each connection line with 5 ml isopropanol prior to installation. 
More extended contact should be avoided, as long-term contact with 
isopropanol may damage the adhesions inside the connecting piece.
9.2.4  Decontaminating the Transport Buffer
t Only use cleaned transport containers for transportation.
t Do not refill the transport container with fresh transport buffer, but use a 
new | cleaned transport container.
t Filter the buffer under sterile conditions (in accordance with GMP).
9.2.5  Decontaminating the Tubing Set
t Clean the tubing set with a cleaning solution (BPT0044A/BPT0044G).