BioPAT
®
Trace | Multi Trace Operating Instructions 69
Initial Installation
6.8 Assignment of Analog Outputs
The measured values remain available for further
use as analog signals at the ports analog 1 and
analog2. When using the ethanol | methanol
method, the signal is output on analog1. The
output analog3 is not assigned.
Measuring
channel
Measuring
channel
Pin 1 [glucose] Pin 2 [lactate]
1 0 (2) – 10 V 1 0 (2) – 10 V
2 GND 2 GND
3 0 (4) – 20 mA 3 0 (4) – 20 mA
(Measurement channel Pin 3 is not assigned).
In the “Parameters” menu (8.3.7.1), the output
format can be switched between 0 – 10 V and
2–10 V or 0 – 20 mA and 4 – 20 mA. The output
areas of the analog outputs are also configured
here.
The output areas of the analog outputs can be
configured as described in Chapter “8.3.7.1
Parameters”, page 90.
The signal at the analog output is changed
whenever there is a new measurement value and
remains unchanged unless:
there is a new measurement value,
the device is rebooted,
“new tubing set” is activated.