EasyManua.ls Logo

Sartorius BioPAT Trace - Page 90

Sartorius BioPAT Trace
137 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
90 BioPAT
®
Trace | Multi Trace Operating Instructions
Operation
8.3.7.1 Parameters
All relevant parameters for measuring operations
can be configured in the “Parameters” sub-item.
Measurement Period
t Enter the desired measuring interval here.
When the value 1 is entered, measurement is
performed every minute or at the maximum
frequency. If one analysis should be per-
formed every hour, then enter 60.
y The frequency can be changed at any time
and becomes immediately effective by
clicking “Ok”.
Autocalibration
y Select yes/no. If you allow autocalibration,
recalibration takes place automatically after
every x number of measurements (see:
measurements between calibrations).
Delay Calibration
t Normally, a calibration is performed after
acertain number of measurements. Longer
interruptions (e.g. measurements every
6 hours for cell cultivations) can lead to very
long periods expiring between the most
recent calibration and the next measurement.
Activating the Delay calibration function
schedules the calibration to be performed
shortly before the next measurement. This
way, an up-to-date calibration is available.

Table of Contents

Related product manuals