EasyManuals Logo

Sartorius BioPAT Trace User Manual

Sartorius BioPAT Trace
137 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #91 background imageLoading...
Page #91 background image
BioPAT
®
Trace | Multi Trace Operating Instructions 91
Operation
Number of Measured Values in Groups
t When using larger measuring intervals, it can
make sense to run several measurements with
maximum frequency in quick sequence at a
given point of time. This way, for example, a
group of 10 measurements may be carried out
every 6 hours. The number of measured
values in the group can be set here. No
average is calculated. Every measured value is
displayed immediately after measurement. If
the time required for the performance of the
group is longer than the measurement period,
then the next group starts with the new
measuring interval.
For the determination of the number of measure-
ments between calibrations, every measurement
in a group is taken into account. This must be
given due consideration when selecting the cali-
bration settings!
Standard Levels (Glucose and Lactate)
t Here, you indicate the concentrations of the
calibration standard solutions that have been
selected and hooked-up for your application.
Measurements Between calibrations
t Here, you can enter how many measurements
should be performed before an automatic
calibration is run. This feature only works,
however, if the autocalibration function has
been allowed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius BioPAT Trace and is the answer not in the manual?

Sartorius BioPAT Trace Specifications

General IconGeneral
BrandSartorius
ModelBioPAT Trace
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals