92 BioPAT
®
Trace | Multi Trace Operating Instructions
Operation
By pressing the “Calibration” button, you can ini-
tiate a calibration at any time. Any changes made
to the automatic calibration data during running
operations are not valid until the next calibration.
Measurements Per Standard
t This function is used to set the number of
measurements per calibration standard. One
minute is required for each measurement of a
calibration value. That means that a normal
calibration with 3 measurements per standard
will take 6 minutes (including rinsing times for
changing the medium).
Analog Out Glucose | Analog Out Lactate
t Here, the output of the signals on the analog
outputs can be selected between 0–20mA
and 4–20mA, and between 0–10V and
2–10V, respectively.
Output Range Glucose | Lactate
t Here, the output ranges of the analog outputs
are set. Proceed as follows: The upper limit of
the measuring range is always indicated. The
lower limit is preset to 0.
For a required measurement range of e.g.
0–10 g/L glucose, glucose is set to 10 (upper
limit) in the output range.
When selecting the voltage range of 0–10V, then
10 volts are equal to 10g/L glucose.