BioPAT
®
Trace | Multi Trace Operating Instructions 125
Malfunctions
Possible cause What should you do?
9. The dialysis membrane was installed in
the probe in the incorrect position
(only for DIALYSIS application).
Correctly install a new
dialysis membrane into
the probe.
10. Three-way stopcock was not properly
adjusted (only for DIALYSIS
application. An external sample was
measured and the three-way stopcock
was not reset).
Adjust the three-way
stopcock to the correct
position.
11. The software versions of the BioPAT
®
Trace | Multi Trace and the PC are not
compatible.
Use suitable software.
If necessary, reinstall the
PC software “trace_mon”
from the website.