EasyManua.ls Logo

Sartorius BioPAT Trace - Page 32

Sartorius BioPAT Trace
137 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
32 BioPAT
®
Trace | Multi Trace Operating Instructions
Design and Function
For the dialysis method, the internal diffusion module is used to calibrate the
BioPAT
®
Trace | Multi Trace. By contrast, the dialysis probe in the bioreactor is
always used for measuring the sample. A reference factor between the two
membranes is then determined once.
As a standard cannot be installed in the bioreactor, the BioPAT
®
Trace | Multi
Trace must be calibrated indirectly by way of the reference factor. During this
process, the initial concentration in the bioreactor is used as the starting con-
centration for the dialysis probe. Using the sample measurement values with
an identical concentration, the BioPAT
®
Trace | Multi Trace calculates the fac-
tor between the diffusion module membrane and the dialysis probe
membrane.
Hence, the calibration procedure for the dialysis method consists of three
steps in total:
Step 1
Recording the calibration curves on the diffusion module with standards 1
and 2.
Step 2
Measuring the external sample on the BioPAT
®
Trace | Multi Trace with the
help of the calibration curves (alternatively: by directly entering the concen-
tration values, if known from the initial weight or by external analysis).
Step 3
Recording the calibration curves of the dialysis probe using the probe.
After measuring the standards for calculating the calibration curves, the
BioPAT
®
Trace | Multi Trace computes a factor from the gradients of each
individual curve. For all measurements carried out with the dialysis probe, this
factor is applied to calculate the true concentration.

Table of Contents

Related product manuals