BioPAT
®
Trace | Multi Trace Operating Instructions 47
Initial Installation
Now place the guidance plates on the valves
such that the pair of tubes are positioned on their
corresponding valve slots.
When inserting the tubing, make sure that the
red, yellow and green color coding is always
located to the right of the valve and matches that
valve’s color coding.
During insertion, hold both sides of the tubing
taut and press the pair down until both tubes are
securely seated in their slots (see figures, below).
While doing this, press the black buttons on the
valve alternately so that you can also insert the
rear tubing deep into the valve slot. Then check
that the tubes are fitted correctly in the valve in
order to avoid measuring errors.