48 BioPAT
®
Trace | Multi Trace Operating Instructions
Initial Installation
After you are done fitting all three valves, push
both guidance plates towards the valve in order
to secure the position of the tubes in the
valveslot.
As the next step, install the two pump cassettes
on the motor shaft and press on the retaining
plate firmly until you hear an audible click
indicating that it has locked in.
CAUTION
Before installing the pump cassettes, check if
themotor shaft is clean (if necessary, wipe with
asoft cloth). Dirt particles or abrasion from
previous applications can adversely affect or
even completely prevent the proper functioning
of the pump.
The lower pump (P2) and the upper pump (P1)
are attached one after the other: