EasyManua.ls Logo

Sartorius BioPAT Trace - Page 28

Sartorius BioPAT Trace
137 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
28 BioPAT
®
Trace | Multi Trace Operating Instructions
Design and Function
Rear view of the device with the following elements:
Figure 8 shows the rear view of the BioPAT
®
Trace | Multi Trace. The device
can be switched on and off by pressing the On/off switch(9). Using the
calibration data, the electrochemical signal of the biosensor is converted
internally to a concentration value. The measured value is output after the
analysis is completed on a PC interfaced to the Ethernet connection(10).
Optionally, the signals can be output via the analog signal outputs(12). The
following ranges are possible: 0..10V, 0..20mA, 4..20mA. The USB interface
(11) and the serial interface (RS232 port) (13) can also be used for data transfer
(currently not implemented).
12a
12b
12c
13
11
10
9
Fig. 8: Connections
Pos. Function Pos. Function
9 On/off switch 12 a | b | c Analog outputs (1–3)
10 Ethernet connection 13 Serial interface
11 USB interface

Table of Contents

Related product manuals