Shimadzu LC-10/20 System 4 Using the control module
Reagent
Fig 40: Method Setup - AS - Pretreatment (Reagent)
Part of the fields (namely Mixing Count, Mixing Volume , Bubbling and
Wait Time) in the Reagent mode has the same meaning as in the Dilution
mode. Different fields are described lower:
Source Vial
Specifies the source vial for the original sample (before reaction). The
automatic calculation uses the same rules as in the Dilution mode.
Reagent Vial 1..3
Specifies vials used for particular reagents. It is possible to select the Not
Used option.
Volume
Specifies the volume of the sample and each reagent used.
- 54 -