Shimadzu LC-10/20 System 4 Using the control module
4.2 Detector
4.2.1 Shimadzu LC-10/20 System Setup - Detector
The Shimadzu LC-10/20 System Setup - Detector tab for SPD detectors
and RF detectors looks the same, although some items may be missing
depending on the type of the detector.
Fig 21: Shimadzu LC-10/20 Detector setup
Base Period
Specifies the base rate of the detector signal acquisition. The value set
here is also limiting the available range of values for the Sampling Rate
as set in the Method Setup - Acquisition dialog - the lower Base Period is,
the higher can the Sampling Rate be. The possible values are influenced
by the checking/unchecking of the Dual Mode item.
Dual Mode
Check to perform the data acquisition on two wavelengths.
Use Recycle Valve
Defines whether the solvent recycling valve is used during the acquisition.
Signal 1 (2) Name
Sets the name of the given detector.
- 30 -