EasyManuals Logo

Shimadzu LC-10 User Manual

Shimadzu LC-10
78 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #40 background imageLoading...
Page #40 background image
Shimadzu LC-10/20 System 4 Using the control module
4.2.2.1 Method Setup - Acquisition - RF Detector
This sub-tab replaces the Detector sub-tab when one of the configured
RF-10A and RF-10Axl detectors is selected in the Select Detector field.
Fig 24: Method Setup - Acquisition - RF Detector
Excitation Wavelength
Sets the excitation wavelength. The possible values are in the range 200
to 900 nm.
Emission Wavelength
Sets the emission wavelength. The possible values are in the range 200
to 900 nm.
Lamp
Sets whether the lamp should be activated prior to the analysis.
Gain
Selects the detector gain amplification value. Possible options are x1, x4
and x16.
Sensitivity
Sets the detector sensitivity by adjusting the voltage on the
photomultiplier.
Response
Defines the detector signal response. Possible values vary for the RF-10A
and RF-10Axl detector.
Sampling Rate
Sets the sampling rate of the detector. The available detector sample rates
are dependent on the Base Period settings on the Shimadzu LC-10/20
- 34 -

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu LC-10 and is the answer not in the manual?

Shimadzu LC-10 Specifications

General IconGeneral
BrandShimadzu
ModelLC-10
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals