Shimadzu LC-10/20 System 4 Using the control module
4.3.2.2 Method Setup - AS - Sampler (SIL-10AF)
This sub-tab replaces the Sampler sub-tab when the SIL-10AF sampler is
configured on the Instrument. The majority of items in the dialog is the
same, the differing items are described below:
Fig 37: Method Setup - AS - Sampler (SIL-10AF)
Excess Volume
Allows to set the excess volume of the sample drawn into the syringe
during each injection. This portion of the sample is not injected, but serves
for better precision of the injection. Allowed values in the Excess Volume
field are between 10 and 100 μl.
Injection Compensation Factor
The value that is used to compensate the actual volume of different
injectors. The specified Injection volume is multiplied by the Injection
Compensation Factor value to produce the real injection volume. Possible
values are 1.00 - 1.30, although usually the value 1.00 is used.
- 50 -