EasyManua.ls Logo

Shimadzu LC-10 - Method Setup - Thermostat - Thermostat

Shimadzu LC-10
78 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Shimadzu LC-10/20 System 4 Using the control module
4.5.2 Method Setup - Thermostat - Thermostat
Fig 50: Method Setup - Thermostat - Thermostat
Enable Thermostat
Enables the temperature control of the thermostat. When unchecked, all
other fields on the tab will be grayed and no thermostat control will be
performed.
Target Temperature
Sets the target temperature. The control module will not switch to the
READY state until this temperature is reached.
Maximum Temperature
Sets the maximal allowed temperature. When the temperature in the
thermostat exceeds the value entered here, the controller will report error
and the analysis will stop.
Initial Valve Position
Sets the initial position of the valves (if there are any installed).
- 66 -