Shimadzu LC-10/20 System 4 Using the control module
4.2.2 Method Setup - Acquisition - Detector
In case of greater number of detectors configured on one Instrument, it is
possible to switch to the desired detector by selecting it in the Select
Detector menu on the upper part of the Method Setup - Acquisition
window.
The chapter describes the SPD detectors, other detector types are
described later in the manual.
Fig 23: Method Setup - Acquisition - Detector
Wavelength Ch.1
Defines the initial wavelength of the first channel.
Wavelength Ch.2
Defines the initial wavelength of the second channel, available only in the
Dual Mode. Any changes made on the tab of one channel will be also
changed on the other channel of the Dual Mode detector.
Lamp
Sets the lamp used at the beginning of the analysis. Available options are
(based on the type of the detector): Off, D2, W, D2+W
Polarity
Selects the initial polarity of the detector.
Response
Defines the level of the digital noise filter. The higher the number set, the
lower the noise level is. On the other hand, some peaks may not be
detected.