Operation at the measuring device
8.5.10 Submenu "MF1 – Srvice"
This submenu is used to parameterize the multifunction output MF1 as
serviceoutput.Youmayactivate(on)ordeactivate(o)severaloptions.
You can get to the menu "Srvice" via the menu path:
Main menu
Menu Pron more
MFx On MF1 Actsta Functn Srvice
Requirements for the display
Menu "more": Option "Yes"
Menu "MFx On": Option "Enable"
Parameter "Functn": Option "Srvice"
Parameter Description
WrnLsr Activating and deactivating warning messages when the
measuring device must be replaced soon because the laser
ages.
Options
On
O
Factory setting
On
WrnLvl Activate or deactivate warning messages when the damping
value is undercut, e.g. at contamination.
Options
On
O
Factory setting
On
WrnTemp Activate or deactivate warning message when the inner
temperature of the measuring device is outside of the
permissible thresholds.
Options
On
O
Factory setting
On
WrnPlb Activate or deactivate warning when the measurement value
is not plausible. Possible reasons may be incorrect measure-
ments, interruption of the light beam, optical interferences or
electrical interferences.
Options
On
O
Factory setting
On