PROFINET IO-interface
86 © SICK AG • Subject to change without notice • 8021823/126P/2018-12-17
Input values
Rel. Adr. Description
0.0 Bit = 1: Plausibility warning
Measured values are impaired. Occurrence of an error is likely.
Type
1 bit
0.1 Bit = 1: Temperature warning
The ambient temperature is only just within the permissible
range. Occurrence of an error is likely.
Type
1 bit
0.2 Bit = 1: Signal level warning
The measured signal level is close to the permissible range.
Occurrence of an error is likely.
Type
1 bit
0.3 Bit = 1: Laser warning
The laser has reached its service life. Occurrence of an error
is likely.
Type
1 bit
0.4 Bit = 1: Plausibility error
No valid measurement value can be calculated.
Type
1 bit
0.5 Bit = 1: Temperature errors
Thedeviceisalsooperatedoutsideofthespeciedtempera-
ture range.
Type
1 bit
0.6 Bit = 1: Signal level error
The measured signal level is too low.
Type
1 bit
0.7 Bit = 1: Laser error
The laser may have reached the end of its service life.
see page 87, chapter 10.3.9, operating hours
Type
1 bit
1.0 Bit=1:Lasero
1 bit
1.1 Bit = 1: Heating on
Type
1 bit
1.2 Bit = 1: MF1 is active.
Type
1 bit