56 FLOWSIC600-XT · Operating Instructions · 8018846/1BHP/V2-1/2021-05 · © SICK Engineering GmbH
Installation
Subject to change without notice
3.3.5 Signal Processing Unit alignment
The Signal Processing Unit can be rotated for the best possible view on the display and safe
cable routing
(Fig. 26). A stop on the housing prevents the Signal Processing Unit from being turned by
more than 330°:
1 Loosen the three screws on the meter body with an SW 3 Allen key.
2 Turn the meter body to the desired position.
3 Retighten the three screws on the meter body neck that were previously loosened
(5Nm).
Fig. 26 Signal Processing Unit alignment
1 2 3