68 FLOWSIC600-XT · Operating Instructions · 8018846/1BHP/V2-1/2021-05 · © SICK Engineering GmbH
Installation
Subject to change without notice
3.4.6 Cable specifications
Power supply 616 V DC (Ex i) / 12 … 24 V DC (Ex d/de)
1)
OZ-BL-CY 2x1.5 mm² is only suitable for the power supply of the FLOWSIC600-XT.
Digital output, current output, encoder, pressure and temperature sensors
2)
RE-2Y (St)Yv with n x 2 x 0.5 mm² (n pairs) is suitable for transmitting the switching signals.
NOTICE: Requirements on cables and installation
Pay attention to the requirements in EN 60079-14 when selecting the
cables and during installation!
FLOWSIC600-XT must be grounded according to EN 60079-14.
Further legal requirements must be observed for use in explosive
atmospheres.
Because of the danger of crosstalk, switching signals and the data signals
of the RS485 should not be routed in a cable with a common shield.
Specification Remark
Type of cable 2 wires
1)
Connect shielding (if present) to
ground terminal
Min./ max. cross-section Ex i:
0.25 mm² / 1 mm²; 1.5 mm² without
wire end
(24 / 18 AWG; 16 AWG without wire
end)
Ex d/de:
0.5 mm² / 2.5 mm² (20 / 12 AWG)
Maximum cable length Depending on loop resistance;
minimum input voltage must be 6V
DC with Ex i and 12V DC with
Ex d/de
Note for Ex i when safety barriers are
used: The cable length is limited to
75 m for ignition group IIC
Cable diameter 6 12 mm Fixing range of the cable glands
Specification Remark
Type of cable Twisted pair, per switching output,
2)
common shield
Connect shield to ground terminal
Min./ max. cross-section 2 x 0.5 mm² / 1 mm² (20-18 AWG) Do not connect unused conductor
pairs and prevent them from
accidental short-circuit
Maximum cable length Loop resistance: ≤ 250 Ohm
Cable diameter 6 12 mm Fixing range of the cable glands

Table of Contents

Other manuals for SICK FLOWSIC600-XT

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SICK FLOWSIC600-XT and is the answer not in the manual?

SICK FLOWSIC600-XT Specifications

General IconGeneral
BrandSICK
ModelFLOWSIC600-XT
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals