MENU NAVIGATION TAD CONTROL UNIT
Fig. 81: Assigning analog output AO4
6.9.3 Setting the PROFIBUS address in “PROFIBUS ID”
The configured address is assigned to the VISIC100SF after a restart when the device is
connected as “slave” in a PROFIBUS-DP system. Submenu item “PROFIBUS ID” serves to
manage the PROFIBUS address. The valid address range is between 0 ... 126.
Arrow buttons: Increase and decrease the digits.
“Right arrow button”: Activate next digit.
Fig. 82: Entering the PROFIBUS address
Status: Maintenance
Enter
Status: Maintenance
Back
AO Mapping 3.3
Enter
Status: Maintenance
/Para/MapAO/AO4
Back
AO4 3.3.4
Save
Configuration 3
Back
/Para/
Enter
Enter
CO
NO
NO
2
Tem p ..
1 AO1
2 AO2
3 AO3
4 AO4
/Para/MapAO/
2 Scale AO
3 AO Mapping
4 PROFIBUS ID
5 Conv. to µg
The password must be entered to set the PROFIBUS address. Password input, “Chang-
ing numerical parameters (password input)”, page 73.
Status: Measuring Status: Measuring
Enter
/Para/
Back
2 Scale AO
3 AO Mapping
4 PROFIBUS ID
5 Conv. to µg
PROFIBUS ID 3.4
/Para/ID/
Configuration 3
Back
Min. : 0
Max. : 126
Save
015
Save
Enter
NOTE: The new address is only active after a system restart.