5.6 Testing digital/analog outputs and gas sensors
Menu item “Test” serves to test the digital/analog outputs and to confirm a gas sensors
check.
5.6.1 Signal test “IO test”
The following signals can be set and/or tested:
● Analog output, factory setting AO1
● Analog output, factory setting AO2
● Analog output, factory setting AO3
● Analog output, factory setting A04
● Relay for maintenance request (“MRq)
● Relay for device malfunction (“Fail”)
5.6.2 Testing the analog output for the K-value with submenu item “AO1”
Fig. 44: Setting and checking the milliampere setting of the analog output for the “AO1” value
Menu item “Test” is only available when menu item “Maint” has been set to active, see
“Activating the setting range via menu item “Maint””, page 53.
The configuration can be changed via the TAD control unit or the device display. AO4 is
only available for TAD control unit with I/O modules. The VISIC has only 3 analog
outputs.
RUN
X
XXX
Set
SET
Maint
Esc
2 s
>
>
...
...
SET
Test
Esc
MEN
4mA
Set
SET
4mA
Set
MEN
20mA
MEN
Filter
Set
MEN
12mA
Set
>
>
>
>
>
>
Set
MEN
IO
Esc
MEN
AO1
Esc
Esc
Esc
Set
Esc
Set
Esc
The factory setting for the K-value is AO1.
Caution: This configuration can be changed by the customer.
The selected current value is set only after the SET button has been pressed.
Submenu item “Filter” is required in connection with the test tool, description, see
“Menu navigation with keypad to submenu item “Filter””, page 92.
The mA value set on the analog output can be reset using “Maint” -> “inactive”. After
30 minutes, the VISIC100SF switches back automatically to measuring mode, see “Acti-
vating the setting range via menu item “Maint””, page 53.