6.8.2 Testing the analog outputs AO2 - AO4
Description, see “Testing the analog output AO1”, page 79.
6.8.3 Testing the “Fault” relay with submenu item “Fault”
Maintenance mode must be activated.
Fig. 73: Enabling the Fault relay
Pressing the “Set” button activates the relay.
It is now possible to check on the relay or in the control station whether the maintenance
relay has been activated.
6.8.4 Testing the “Maintenance Request” relay with submenu item “Maintenance Req.
Maintenance mode must be activated.
The relay for maintenance request is set the same as the relay for malfunction. Procedure,
see “Testing the “Fault” relay with submenu item “Fault””, page 80.
6.9 Performing settings on the device with menu item “Configuration”
The following settings can be performed via menu item “Configuration”:
Select language (7 languages), see “Selecting the menu language”, page 72.
Scale AO
AO Mapping
PROFIBUS ID
k/μg conversion
Activate/deactivate temperature sensor
Status: Maintenance
/Diag/IOT/
Back Enter
Status: Maintenance
/Diag/IOT/Fault/
Back
Fault 2.5.6
Set
3 AO3
4 AO4
5 Maintenance request
6 Fault
IO Test
1 activate
2 deactivate
Enter
NOTE: The maintenance signal must be activated to change settings.
Set the maintenance signal in the menu, see “Activating Maintenance mode”,
page 74 or
password prompt before setting the values.