8017171/115Z/V3-0/2019-01/ SICK OPERATING INSTRUCTIONS | VISIC100SF
Subject to change without notice
INSTALLATION
3.7.3.4 Modbus-RTU Read Coil (0x01)
Function code “Read Coil (0x01)” serves to read out all malfunction and maintenance
request messages from VISIC100SF.
Table 11: Read Coil Modbus-RTU
Example:
Read coil number 200 from slave (ID 101):
TX-> <65 01 00 C8 00 01 74 10>
RX-> <65 01 01 00 4E B8>
Maintenance request Vis = false
3.7.4 PROFIBUS DP-V0 (optional)
The PROFIBUS module belongs to the VISIC100SF when configured at the same time
during ordering. The VISIC100SF is integrated in the bus via a restart after wiring.
3.7.4.1 PROFIBUS addressing
The PROFIBUS-DP address of the device can be managed via the keypad.
For further information, see Section “Menu”, see “Setting the PROFIBUS address in “PB
ID””, page 57.
Coil number Designation
200 Contaminated optics
201 Limit of CO operating hours reached
202 Limit of NO operating hours reached
203 Maintenance request of external temperature sensor
204-206 Reserved
207 Limit of NO
2
operating hours reached
208-215 Reserved
216 Error Vis
217 Error CO sensor
218 Error NO sensor
219 Error EEPROM
220 Error heating
221 Error 4 ... 20 mA interface
222 Error FPGA
223 Error CPU
224 Error program flow
225 Error enclosure cover
226 Error NO
2
cell
226-229 Reserved
230 Maintenance active
231 Reserved
The device must be restarted after the address is changed.
Press “Reset” to restart, see “Measuring unit - circuit board with display and keypad”,
page 17.