SPECIFICATIONS
10.3 Technical Data
VISIC100SF
Measured variables
Visibility (K-value)
Gas concentration CO/NO/NO
2
(optional)
Measuring principles
Scattered light forwards (K-value)
Gas sensor (CO/NO/NO
2
)
Measuring ranges
Visibility (K-value): 0 ... 15 /km
CO: 0 ... 300 ppm
NO: 0 ... 100 ppm
NO
2
: 0 ... 5 ppm
Optional temperature measurement -30 .. +70 °C
Setting time t
90
60 s
Accuracy
CO: 3% of full-scale value
NO: 3% of upper measuring range value
NO
2
: 2% of the upper measuring range value
Resolution
Visibility (K-value): 0.001 /km
CO: 0.5 ppm
NO: 0.5 ppm
NO
2
: 0.05 ppm
Repeat accuracy
Visibility (K-value): 2%
Ambient temperature
-20 ... +55 °C
Storage temperature
Measuring device without gas sensor:
–30 +85 °C
CO-/NO-/NO
2
sensors: +5 … +20 °C
Ambient pressure 860 1,080 hPa
Ambient humidity 10% ... 100% RF, non-condensing
Electrical safety CE
Control functions
Contamination monitoring of glass pane
Drift and plausibility check
Automatic self-test
Function monitoring of optional heating
System components
Basic variants
Measuring unit with wall enclosure and cover
Optional:
Terminal box
TAD control unit
Gas sensors: CO, NO and NO
2
measurement
Heating
Scope of delivery
Exact device specifications and performance data of the
product can deviate and depend on the respective application
and customer specification.
Protection class IP 6K9K
Analog outputs
3 outputs: 4 ... 20 mA, electrically isolated (max. load
500 Ohm)
Digital outputs 2 outputs: 48 V DC, 0.5 A, 24 W
Interfaces 2 x RS-485
Bus protocol
Integrated: Modbus-RTU
Optional: PROFIBUS DP-V0