CLEARING MALFUNCTIONS
9Clearing Malfunctions
9.1 Description of device errors
The VISIC100SF switches immediately to Malfunction when a device error occurs. In
operating state Malfunction, the malfunction relay opens and the three analog interfaces
signal 1 mA. The digital interfaces (PROFIBUS and Modbus) have a measured value status
which switches to “Bad” when an error occurs. The following Table shows the error codes
displayed for possible device errors.
Table 20: Device error coding
Information on calling up error codes on the display, see “Calling up maintenance
request and malfunction messages with menu item “Status””, page 53.
Code Bit Description Cause Notes for service
F_000 0 VIS error
Contam. Optic.
LED Fault.
VIS>Spec.
Clean and restart device.
Exchange measuring unit. (via SICK Customer
Service).
F_001 1 CO sensor
CO Sensor Fault,
Warmup Sensor.
Wait for heating up time to complete.
Restart.
Exchange gas sensor.
F_002 2 NO Sensor
NO Sensor Fault,
Warmup Sensor.
Wait for heating up time to complete.
Restart.
Exchange gas sensor.
F_003 3 EEPROM EEPROM Data Inconsistent.
Restart. If the error is still present after a
restart, call SICK Customer Service or return the
device, and specify the error code.
F_004 4 Heating
Enclosure cover not fitted because
power supply interrupted -> no
heating error.
Heating Current Out Of Spec
Fit the enclosure cover.
Restart. If the error is still present, call SICK
Customer Service.
Exchange cover.
F_005 5
Erroneous
function of
analog inter-
faces
Electronic Fault.
Restart. If the error is still present after a
restart, call SICK Customer Service or return the
device, and specify the error code.
F_006 6 FPGA
FPGA Fault,
ADC Overload.
Restart. If the error is still present after a
restart, call SICK Customer Service or return the
device, and specify the error code.
F_007 7 CPU
RAM Test Fault.
Flash Test Fault.
Register Test Fault.
Call SICK Customer Service or return the device,
and specify the error code.
F_008 8 Program flow Program Flow Fault.
Restart. If the error is still present after a
restart, call SICK Customer Service or return the
device, and specify the error code.
F_009 9
Enclosure
error
Enclosure Cover not Fitted.
Fit the enclosure cover.
F_010 10 NO
2
Sensor
NO
2
Sensor Fault,
Warmup Sensor.
Wait for heating up time to complete.
Restart.
Exchange gas sensor.
F_014 14 Maintenance Maintenance Active.
Deactivate maintenance on device, see “Acti-
vating maintenance in menu item “Maint””,
page 53.
The status is always shown in plain text on the basis display.