CLEARING MALFUNCTIONS
9.2 Description of maintenance requests
Table 21: Description of maintenance requests
9.3 Display of error states on the TAD control unit
9.4 Further error causes
Data interruption through VISIC100SF self-test
Self-tests are performed every four hours for RAM/Flash and CPU registers. Short
interruptions (between 8 μs and 140 ms) in communication to the Modbus-RTU/TAD
control unit interface are therefore possible and can lead to transfer errors/timeouts on the
Master.
Code Bit Description
Coding
Maintenance requests
Notes for service
MRq_000 0
VIS
measurement
Contamination Limit 1 Reached
Clean enclosure and optics.
Restart.
MRq_001 1 CO Sensor Maintenance of CO sensor required
Exchange gas sensor.
Calibrate gas sensor if necessary.
MRq_002 2 NO Sensor Maintenance of NO sensor required
MRq_003 3 Temp Temperature Sensor Fault.
Exchange temperature sensor
MRq_004 4 DO Module Communication Fault DO-Modul.
Calibrate DO module.
MRq_005 5 AO Module Communication Fault AO-Module
Exchange AO module.
MRq_006 6
TAD control
unit
Communication Fault TAD control
unit
Exchange TAD control unit.
MRq_007 7 NO
2
Sensor
Maintenance of NO
2
sensor
required
Exchange gas sensor and calibrate anew if
necessary.
Indication Actions
“POWER” off
▸ Check main power supply (external main power switch,
mains fuses)
“FAILURE” on
▸ Check messages.
Measured values blink
“MAINTENANCE REQUEST” on ▸ Check in menu item Diagnosis
which maintenance request exists.
Measured values implausible ▸ Check whether it is possible for the measured values to
reach these values in the current situation.
▸ Check device for contamination.