MAINTENANCE
Special case: Actual value is to be output via the analog output for reading
Submenu item “Filter” must be activated via the keypad on the VISIC100SF to enable
transfer of the values to the central computer in the control station.
By activating submenu item “Filter”, the actual value is not only shown on the display but
also output on the analog output.
Fig. 91: Menu navigation with keypad to submenu item “Filter”
Afterwards, perform test as described above.
8.3.1.4 Gas sensor maintenance
Exchanging the gas sensors
1 Switch the VISIC100SF off.
2 Disconnect the data cable from the circuit board.
3 Use an SW4 Allen key to loosen the gas sensor screw fitting on the underside of the
VISIC100SF enclosure.
4 Unscrew the loose sensor by hand.
5 Gas sensor disposal, see “Disposal”, page 87.
6 Start-up of new sensor, see “Installing the TAD control unit (optional)”, page 25.
8.3.1.5 Readjustment of gas sensors
Gas sensors can be readjusted with the optional gas calibration case PN: 2080119.
RUN
X
XXX
Set
SET
Maint
Esc
2 s
>
>
...
...
SET
Test
Esc
MEN
4mA
Set
SET
4mA
Set
MEN
20mA
MEN
Filter
Set
MEN
12mA
Set
>
>
>
>
>
>
Set
MEN
IO
Esc
MEN
AO1
Esc
Esc
Esc
Set
Esc
Set
Esc
NOTE: Check correct wiring of malfunction relay
The test value is shown as actual value if the malfunction relay is not connected and
causes an incorrect ventilator control.
WARNING: Health risk through sulfuric acid
The gas sensor contains sulfuric acid.
▸ When taking the gas sensors out, avoid any damage by a pointed or sharp object.
Package the gas sensors carefully and securely when damaged and dispose of as
hazardous waste.