156 DCA Vantage Operator’s Guide
Customizing the System
Setting the Albumin User Correction
Use the Albumin User Correction screen to adjust the slope and the offset for
Albumin values.
NOTE: Siemens recommends that you do not use a user defined correlation
when you perform external proficiency testing. This can impact the results.
1. At the Home screen, select Menu.
2. Select
System Settings from the Menu screen.
3. At the System Settings menu, select
Additional Settings.
4. At the Additional Settings menu, select
Patient Tests.
5. Select
Albumin User Correction.
6. Change the slope, if needed:
a. Select
Edit.
The Albumin Slope screen displays.
b. Enter the slope.
c. Select Enter.
The Albumin User Correction screen displays.
7. Change the offset, if needed:
a. Select
Edit.
The Albumin Offset screen displays.
b. Enter the offset.
c. Select Enter.
The Albumin User Correction Screen displays.
To reset the settings to the default values:
a. Select
Reset Default Values.
This resets the slope to 1.000 and the offset to 0.0.
8. After completing the changes, select Back.
The Patient Tests screen displays.
Setting the Creatinine User Correction
Use the Creatinine User Correction screen to adjust the slope and the offset
for creatinine values.
1. At the Home screen, select
Menu.
2. Select
System Settings, from the Menu screen.
3. At the System Settings menu, select
Additional Settings.
4. At the Additional Settings menu, select
Patient Tests.
5. Select
Creatinine User Correction.

Table of Contents