DCA Vantage Operator’s Guide 157
Customizing the System
Customizing the System
6. Change the slope, if needed:
a. Select Edit.
The Creatinine Slope screen displays.
b. Enter the slope.
c. Select
Enter.
The Creatinine User Correction screen displays.
7. Change the offset, if needed:
a. Select Edit.
The Creatinine Offset screen displays.
b. Enter the offset.
c. Select
Enter.
The Creatinine User Correction Screen displays.
8. To reset the settings to the default values, select Reset Default Values.
This resets the slope to 1.000 and the offset to 0.0.
9. Select
Back.
The Patient Tests screen displays.
Setting the GFR Calculation
Use the GFR Calculation screen to turn the GFR Calculation on or off, and to
select the unit of measurement for plasma creatinine.
1. At the Home screen, select Menu.
2. Select
System Settings from the Menu screen.
3. At the System Settings menu, select
Additional Settings.
4. At the Additional Settings menu, select
Patient Tests.
5. Select
GFR Calculation.
6. Select one of the following options under GFR Calculation:
Enabled
Disabled
7. Select one of the following units under Creatinine Units:
mg/dL
•mmol/L
8. To save the GFR Calculation settings, select Save.
Setting the Patient Recall List Options
Use the Patient Recall List Options screen to set the search criteria for patient
results, and how the results are displayed. You can search by either patient ID
or last name.