168 DCA Vantage Operator’s Guide
Safety Information
Disinfect tools and other items that have been near any part of the system
sample path or waste area with 10% v/v bleach.
Do not eat, drink, smoke, or apply cosmetics or contact lenses while in the
laboratory.
Do not mouth pipet any liquid, including water.
Do not place tools or any other items in your mouth.
Do not use the biohazard sink for personal cleaning such as rinsing coffee
cups or washing hands.
Do not recap, purposely bend, cut, break, remove from disposable
syringes, or otherwise manipulate needles by hand. Needlestick injuries
may result.
References
1. Centers for Disease Control. Update: Universal precautions for prevention
of transmission of human immunodeficiency virus, hepatitis B virus and
other bloodborne pathogens in healthcare settings. 1988. MMWR, 37:377–
382, 387, 388.
2. Clinical and Laboratory Standards Institute (formerly NCCLS). Protection
of Laboratory Workers from Occupationally Acquired Infections; Approved
Guideline - Third Edition. Wayne, PA: Clinical and Laboratory Standards
Institute; 2005. CLSI Document M29-A3. [ISBN 1-56238-567-4].
3. Federal Occupational Safety and Health Administration. Bloodborne
Pathogens Standard. 29 CFR 1910. 1030.
External Barcode Reader Beam Safety
Never look directly at the beam of a handheld barcode reader.
Never point the scanner at another person.
Do not look at the reflection of the beam from a shiny surface.
Electromagnetic Emissions Safety
When operating the DCA Vantage system, observe the following precautions:
Do not use this system in close proximity to sources of strong
electromagnetic radiation (for example, unshielded intentional RF
sources), as these may interfere with the proper operation.
Evaluate the electromagnetic environment prior to operating the system.
The system complies with the emission and immunity requirements described
in IEC 61326-1 and IEC 61326-2-6 series.