DCA Vantage Operator’s Guide 73
Operating the System
Result Details
The Microalbumin/Creatinine test displays 3 values:
• Albumin concentration
• Creatinine concentration
• Albumin/Creatinine ratio
Both the albumin and creatinine tests are linear throughout their ranges.
NOTE: If there is an asterisk next to the result percent, this indicates that the
results were user corrected.
The following table displays the reporting range for each test.
NOTE: If creatinine units are mg/dL, then the Albumin/Creatinine (A/C) ratio is
reported as mg/g. If the creatinine units are mmol/L, then the Albumin/
Creatinine (A/C) ratio is reported as mg/mmol.
An albumin or creatinine value preceded by a less than sign (<) in the display,
indicates that the concentration is below the lower limit of the test, that is
< 5 mg/L for albumin or < 15 mg/dL for creatinine.
An albumin or creatinine value preceded by a greater than sign (>) in the
display, indicates that the concentration is above the upper reporting limit of
the test, that is > 300 mg/L for albumin and > 500 mg/dL for creatinine.
An Albumin/Creatinine ratio preceded by a less than (<) sign or greater
than (>) sign or (---) in the display, indicates that either the albumin and/or
creatinine value is above or below the Reporting Range. In certain cases, no
ratio is reported (---).
Albumin/Creatinine ratio The Albumin/Creatinine ratio results
display in mg/g units if Creatinine units
are mg/dL, or in mg/mmol if creatinine
units are mmol/L.
Albumin user correction slope and
offset
The Albumin user correction slope and
offset display.
Creatinine user correction slope
and offset
The Creatinine user correction slope
and offset display.
Test Reporting Range
Albumin 5–300 mg/L
Creatinine 15–500 mg/dL (1.3–44.2 mmol/L)
Albumin/Creatinine ratio 1–2000 mg/g (0.11 to 226 mg/mmol)
Result Description