ID/
Icons
Diagnostic Action Comments
62 Fluid temp. below limit Increase the fluid temperature. If the failure con‐
tinues then contact your local Siemens repre‐
sentative
63 Fluid temp. above limit Reduce the fluid temperature. If the failure con‐
tinues then contact your local Siemens repre‐
sentative
64 Frame temp. below limit Increase fluid temperature and check that ambi‐
ent temperature is within specified limits. If the
failure continues then contact your local Sie‐
mens representative
65 Frame temp. above limit Reduce fluid temperature and check that ambi‐
ent temperature is within specified limits. If the
failure continues then contact your local Sie‐
mens representative
66 Standard Deviation' above limit Measurement continues with values from last
successful zero point adjustment. Improve con‐
ditions for automatic zero point adjustment and
repeat adjustment.
67 Zero Point Offset' above limit Measurement continues with values from last
successful zero point adjustment. Improve con‐
ditions for automatic zero point adjustment and
repeat adjustment.
68 Zero point adjustment failed Measurement continues with values from last
successful zero point adjustment. Improve con‐
ditions for automatic zero point adjustment and
repeat adjustment.
69 Empty Tube Limit' exceeded Make sure that the sensor is filled with liquid and
that the liquid density is within the specified
'Empty Tube Limit'
70 Too little fluid in tube Make sure that the sensor is filled with liquid
Diagnosing and troubleshooting
10.2 Fault codes and corrective actions
MASS 2100 & FC300 (FCT030)
192 Operating Instructions, 06/2017, A5E39789040-AA